Honeywell MIDAS Sensor Cartridge for Ozone (O3) 0.036-0.4 ppm

$414.00

Description

Honeywell MIDAS Sensor Cartridge for Ozone (O3) 0.036-0.4 ppm

Several Honeywell MIDAS® cartridges can be configured to read a number of different gases based on a common chemical family. Actual details for the sensor detection range are included in the sensor cartridge specification sheet and the MIDAS® technical manual. MIDAS® sensor cartridges can be ordered as part of a Kit with the universal MIDAS® transmitter, or as a replacement part. The detectable range is listed. With the exception of O2 (where ambient concentration is typically 20.9%), the sensors will display zero when no gas is present. All Cartridges with the format MIDAS-E-XXX are supplied with a 2 year (24 month) warranty.
Gas MeasuredOzone (O3)
Cartridge Part NumberMIDAS-E-O3X2 year extended warranty
Sensor Technology3 electrode electrochemical cell
Measuring Range (ppm)O30 – 0.4 ppm
Minimum Alarm 1 Set Point0.050 ppm
Repeatability< ± 5 % of measured value
Linearity< ± 5 % of measured value
Response Time t62.5< 60 seconds
Sensor Cartridge Life Expectancy≥ 24 months under typical application conditions
Operating Temperature0°C to +40°C (32°F to 104°F)
Effect of TemperatureZero< ± 0.0018 ppm / °C
Sensitivity< ± 0.5% /°C
Operating Humidity (continuous)15 – 90% rH
Effect of HumidityZeroAbrupt changes will cause a short term drift
Sensitivity< ± 1 % of measured value / % rH
Operating Pressure90 – 110kPa
Effect of PositionNo effect in typical application
Long Term DriftZeroNo Drift
Sensitivity< 5 % of measured value / 6 months
Calibration GasOzone (O3)
Challenge Gas (Bump Test)Chlorine (Cl2)
Warm Up Time< 10 minutes
Storage Temperature+5°C to +25°C (+41°F to +77°F)
Cross Sensitivities

Each Midas sensor is potentially cross sensitive to other gases and this may cause a gas reading when exposed to other gases than those originally designated. The table below presents typical readings that will be observed when a new sensor cartridge is exposed to the cross sensitive gas (or a mixture of gases containing the cross sensitive species).
Gas / VaporChemical FormulaConcentration Applied (ppm)Reading (ppm HF)
AmmoniaNH3100-3
ArsineAsH30.20
Carbon DioxideCO250000
Carbon MonoxideCO1000
ChlorineCl211.2
Chlorine DioxideClO211.5
Chlorine TrifluorideClF311 (Theoretical)
FluorineF20.10.1
HydrazineN2H43-3
HydrogenH230000
Hydrogen SulfideH2S201.6
Nitric OxideNO1001
NitrogenN2100 %0
Nitrogen DioxideNO2106
Sulfur DioxideSO220-0.2

Additional information

Weight 1 lbs