Honeywell Midas-M Oxygen (O2) Sensor, 0-25% v/v, 3 years

$633.00

Description

Honeywell Midas-M Oxygen (O2) Sensor, 0-25% v/v, 3 years

The Midas-M Sensor Cartridge enables the detection of a wide range of asphyxiant, toxic, flammable, pyrophoric, corrosive, and oxidizer gases. The plug & play Sensor Cartridge is located inside the detector’s chassis.
Gas MeasuredOxygen (O2)
Cartridge Part NumberMMS-D2
Sensor Technology3 electrode electrochemical cell
Measuring RangeO20% v/v to 25% v/v
Default Alarm 123.5% v/v (rising)
Default Alarm 219.5% v/v (falling)
Accuracy<±0.2% v/v at 20.9% v/v O2
Response Time (t62.5)Typical 11 seconds
Sensor Cartridge Life Expectancy36 months under typical application conditions
Operating Temperature0ºC to 40ºC (32ºF to 104ºF)
Effect of TemperatureZeroTBA
Sensitivity<±0.2% of measured value/ºC
Operating Humidity (continuous)15% RH to 90% RH
Effect of HumidityZeroFollows actual concentration of O2 present
Sensitivity(eg. 20.9% v/v @ 30% RH, 20.04% v/v @ 99% RH/40°C)
Operating Pressure70 kPa to 110 kPa
Effect of PositionNo effect in typical application
Long Term DriftZeroNo drift
Sensitivity<5% signal loss over operating life
Calibration GasOxygen (O2 20.9% v/v)
Challenge Gas (Bump Test)Air mixture
Warm Up Time< 30 minutes
Storage Temperature5°C to 25°C (41°F to 77°F)
Cross Sensitivities

Each Midas-M sensor is potentially cross sensitive to other gases and this may cause a gas reading when exposed to other gases than those originally designated. The table below presents typical readings that will be observed when a new sensor cartridge is exposed to the cross sensitive gas (or a mixture of gases containing the cross sensitive species).
Gas / VaporChemical FormulaConcentration AppliedReading (%v/v O2)
Carbon DioxideCO2Enhance O2 reading by 0.3% / % CO2
HydrogenH2100% v/v-9
MethaneCH4100% v/vNo response
Nitrogen DioxideNO225 ppm in airNo response

Additional information

Weight 1 lbs