Description
35-3001A-08-02 IR Hydrocarbon 100% LEL / Carbon Monoxide 0-300 ppm with transmitter
The 35-3001 series is a compact sample draw detector assembly with a built-in pump that is powered from 24VDC. This unit features a NEMA 4X enclosure, making it dust, water and corrosion resistant. It also features a low flow alarm which warns of any obstructions or restrictions in the flow system. Combustibles, oxygen, and carbon dioxide versions are available with or without a 4-20 mA transmitter. H2S and CO versions both have 4-20mA outputs. The 35-3001 is also available in many dual sensor configurations.
The 35-3001 is capable of single person calibrations and remote sampling at up to 5,000 ft. from a controller, and interfacing to any RKI or third party control system (utilizing a 4-20 mA feedback signal).
- Available in LEL, O2, H2S, CO, CO2, H2 (ppm or LEL)
- Multigas versions available with sensors in one enclosure
- Internal sample pump with 100’ range
- Flowmeter and LED’s for operational status
- Low flow indication
- NEMA 4X enclosure
- Operates on 24VDC input
- Optional 4-20mA output
- Petrochemical plants
- Refineries
- Water & wastewater treatment plants
- Pulp & paper mills
- Gas, telephone, & electric utilities
- Parking garages
- Manufacturing facilities
- Steel
- Automotive
- HVAC
Target Gases and Detection Range | Combustible Gas (methane calibration standard): | 0 - 100% LEL |
Carbon Monoxide: | 0 - 300 ppm | |
Input Power | 24 VDC Nominal (18.5 VDC - 30 VDC) | |
Construction (housing) | Fiberglass/polyester (NEMA 4X) | |
Dimensions | Height | 8.5 in. |
Width | 6.5 in. | |
Depth | 4.25 in. | |
Weight | 4.5 lbs. | |
Sampling Method | Sample-draw | |
Sample Flow | 3.0 SCFH typical, with no inlet or exhaust line | |
Flow to Sensor | 1.0 SCFH (nominal) | |
Low Flow Setpoint | 0.6 ± 0.1 SCFH | |
Maximum Recommended Inlet/Exhaust Line Length for 1/4” O.D. x 1/8” I.D. Tubing | Inlet | Exhaust |
50 feet | 0 feet | |
Maximum Recommended Inlet/Exhaust Line Length for 1/4” O.D. x 0.170” I.D. Tubing | Inlet | Exhaust |
50 feet | 50 feet | |
75 feet | 0 feet | |
Response Time | 90% in 30 seconds | |
Accuracy | Combustible Gas: | ± 5% of reading or ± 2% LEL (whichever is greater) |
Carbon Monoxide: | ± 5% of reading or ± 5 ppm CO (whichever is greater) |