Sensepoint
Sensepoint with Junction Box
The Sensepoint range of flammable, toxic and oxygen gas detectors offer users a low cost solution to their gas monitoring needs, indoors or out and even in potentially explosive atmospheres (ATEX approval only). Supplied with a pre-wired junction box, the detector can be installed using conduit or armored cable and glands depending on individual site standards. The junction box includes a hinged lid to ensure both hands are free to complete electrical connections. Threaded M20 as well as 3/4" and 1/2” NPT clearance entries are included as standard, each one fitted with an appropriate seal, so enabling the user to select the most convenient entry size to use.
Fit For Purpose
• Flammable, toxic and Oxygen versions available
• High performance, low cost
• Suitable for new and retrofit applications
Proven Sensor Technology
• Surecell electrochemical sensors
• Poison resistant catalytic bead
• Fast and reliable
• Typical 2 to 5 years operational life
ATEX Certified
• Complies with the latest European standards
• Hazardous area approval and performance
• IP65 / 66 for use in harsh environments
• Factory set operation
• Simple to replace
• Minimal training required
Range of Accessories
• Weatherproof cap
• Flow adaptor
• Duct mounting kit
• Collecting cone
• Flammable, toxic and Oxygen versions available
• High performance, low cost
• Suitable for new and retrofit applications
Proven Sensor Technology
• Surecell electrochemical sensors
• Poison resistant catalytic bead
• Fast and reliable
• Typical 2 to 5 years operational life
ATEX Certified
• Complies with the latest European standards
• Hazardous area approval and performance
• IP65 / 66 for use in harsh environments
• Factory set operation
• Simple to replace
• Minimal training required
Range of Accessories
• Weatherproof cap
• Flow adaptor
• Duct mounting kit
• Collecting cone
General specification | ||||||
---|---|---|---|---|---|---|
Use | Cost-effective, fixed point gas detector for the protection of personnel and plant from flammable, toxic and Oxygen gas hazards | |||||
Detectable gases | ||||||
Gas | Measuring range | Operating temperature | ||||
Identifier | Name | 1 | 2 | 3 | Min | Max |
F | Flammable | 0-100% LEL | -40°C / -40°F | 50°C / 122°F | ||
O | Oxygen | 0-25% VOL | -15°C / 5°F | 40°C / 104°F | ||
C | Carbon Monoxide | 0-200 ppm | 0-500 ppm | -20°C / -4°F | 50°C / 122°F | |
H | Hydrogen Sulfide | 0-50 ppm | 0-20 ppm | 0-100 ppm | -25°C / -13°F | 50°C / 122°F |
G | Hydrogen | 0-1000 ppm | -5°C / 23°F | 40°C / 104°F | ||
L | Chlorine | 0-5 ppm | 0-15 ppm | -20°C / -4°F | 50°C / 122°F | |
S | Sulfur Dioxide | 0-15 ppm | 0-50 ppm | -15°C / 5°F | 40°C / 104°F | |
N | Nitrogen Dioxide | 0-10 ppm | -15°C / 5°F | 40°C / 104°F | ||
A | Ammonia | 0-50 ppm | 0-1000 ppm | -20°C / -4°F | 40°C / 104°F | |
Construction | ||||||
Electrical | Flammable (0-100% LEL) | Toxic & Oxygen | Junction Box | |||
3 wire mV bridge | 2 wire loop powered 4-20mA (plus screen) | Earth continuity plate | ||||
2.9-3.5V, 0.7W | 16-30 Vdc, 0.9W | BK4 terminal block (4 way | ||||
200mA constant current | 4 x 0.5mm2 (20AWG) to 2.5mm2 (14AWG) | |||||
1 x M20, 1 x ¾” NPT, 1 x ½” NPT | ||||||
Material | Polyphenylene Sulfide (PPS) | Polyphenylene Sulfide (PPS) | Fire retardant GFP | |||
Weight | 190g | 185g (205g Oxygen) | 600g | |||
Certification | Ex II 2GD | Ex II 2 GD Ex d ia IIC T4 Gb Tamb -40ºC to +65ºC | Ex II 2GD | |||
Ex d IIC T6 Gb | Ex tb IIIC A21 IP67 T135ºC Db | Ex e IIC T6 Gb Tamb -40ºC to +50ºC | ||||
Ex tb IIIC A21 Db IP67 T85ºC | Baseefa08ATEX0263X | Ex t IIIC T85ºC Db IP66 | ||||
(Tamb -40ºC to +55ºC) or T100ºC | IECEx BAS08.0070X | Sira 06ATEX3185 | ||||
(Tamb -40ºC to +70ºC) or T135ºC | ||||||
(Tamb -40ºC to +80ºC) | ||||||
Baseefa08ATEX0265X | ||||||
IECEx BAS08.0068X | ||||||
Applicable standards | This product complies with the relevant CE standards including: EMC to EN50270, Flammable to EN60079-29-1, Toxic to EN445544 (H2S only) | |||||
Environmental | ||||||
IP rating | IP65 standard, IP66 with weather protection in accordance with EN60529:1992 | |||||
Operating temperature | Dependent upon gas type and certification | |||||
Operating humidity range | Continuous: | 20-90% RH (non condensing) | ||||
Intermittent: | 10-99% RH (non condensing) | |||||
Operating pressure | 90-110kPa | |||||
Storage conditions | Instrument -40°C to +50°C (-40°F to +122°F) | |||||
Ordering Information | ||||||
Standard Supply | Each detector is supplied complete with junction box, multiple cable entries, earth continuity plate, hinged lid and instruction leaflet, all sealed in a plastic bag inside a shipping carton | |||||
Shipping Details | Shipping carton dimensions: | Length: | 6.3 in. (160 mm) | |||
Width: | 8.5 in. (215 mm) | |||||
Depth: | 3.1 in. (80 mm) | |||||
Optional Accessories | 02000-A-1645 | Calibration gas flow housing (for flammable & toxic) | ||||
00780-A-0160 | Right angle mounting bracket for ceiling mounting | |||||
02000-A-1640 | Weather Housing for use with flammable detector including remote gassing nozzle | |||||
02000-A-1635 | Weather Housing for use with toxic/Oxygen detector including remote gassing nozzle | |||||
02000-A-1642 | Collecting cone for use with flammable detector | |||||
Calibration Equipment | Contact Honeywell Analytics Representative |
Gas & Range | Flammable 0-100% LEL | Flammable PPM (Typically 0-10% LEL) | Hydrogen Sulfide (H2S) 20/50/100 ppm | Carbon Monoxide (CO) 100/200/500 ppm | Ammonia (NH3) 50/1000 ppm | Chlorine (Cl2) 5/15 ppm | Oxygen (O2) 25% V / V | Hydrogen (H2) 1000/10000 ppm | Sulfur Dioxide (SO2) 15/50ppm | Nitric Oxide (NO) 100ppm | Nitrogen Dioxide (NO2) 10ppm |
---|---|---|---|---|---|---|---|---|---|---|---|
Operating Temp | -40 to +80˚C | -30 to +65˚C | -25 to +50˚C | -20 to +50˚C | -20 to +40˚C | -20 to +50˚C | -15 to +40˚C | -15 to +40˚C | -15 to +40˚C | -15 to +40˚C | -15 to +40˚C |
Operating Humidity | Continuous: | 20 to 90% RH (non-condensing) | |||||||||
Intermittent: | 10 to 99% RH (non-condensing) | ||||||||||
Storage Conditions | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH | 0 to 25˚C; 30 to 70% RH |
Operating Pressure | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa | 90 to 110kPa |
Warm-up Time | < 10 minutes | < 20 minutes | < 3 minutes | < 3 minutes | < 3 minutes | < 5 minutes | < 5 minutes | < 3 minutes | < 3 minutes | 12 hours | 1 hour |
Response Time | |||||||||||
T20 | – | – | < 6 secs | < 5 secs | – | < 10 secs | < 3 secs | – | – | < 3 secs | < 15 secs |
T50 | < 5 secs | < 5 secs | < 15 secs | < 12 secs | < 10 secs | < 20 secs | < 5 secs | < 10 secs | < 10 secs | < 5 secs | < 30 secs |
T90 | < 10 secs | < 10 secs | < 40 secs | < 30 secs | < 65 secs | < 105 secs | < 10 secs | < 45 secs | < 90 secs | < 30 secs | < 60 secs |
Supply | 200mA; 0.7W maximum | 18 to 30VDC; 0.9W maximum | |||||||||
Signal Output | mV bridge | Loop powered 4-20mA | |||||||||
Linearity | < 5% fsd or ±10% of reading | < 5% fsd | |||||||||
Operating Life | 5 years (typical) | 2 years (typical) | 18 months (typical) | 1 year (typical) | 18 months (typical) | ||||||
Storage Life | Typically, over 5 years when in original sealed packaging | 6 months | 6 months | 6 months | 6 months | 6 months | 6 months | 6 months | 6 months | 6 months | |
Stability (zero) | |||||||||||
with time | < ±3% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | < ±2% fsd / yr | |
with temperature | < ±1% fsd2 | < ±3% fsd | < ±5% fsd | < ±6% fsd | < ±2% fsd | < ±2% fsd | < ±5% fsd | < ±5% fsd | < ±5% fsd | ||
with humidity | < ±1% fsd | < ±10% fsd | < ±2% fsd | < ±10% fsd | < ±2% fsd | < ±2% fsd | < ±10% fsd | < ±10% fsd | < ±10% fsd | ||
Stability (span) | |||||||||||
with time | < ±3% fsd / yr | < ±2% fsd / yr | < ±4% fsd / yr | < ±20% fsd / yr | < 10% fsd / yr | < ±20% fsd / yr | < ±2% fsd / yr | ||||
with temperature | < ±3% fsd | < ±8% fsd | < ±25% fsd | < ±20% fsd | < ±20% fsd | < 40% fsd | < ±10% fsd | < ±60% fsd | < ±15% fsd | < ±25% fsd | |
with humidity | < ±2% fsd | < ±10% fsd | < ±2% fsd | < ±10% fsd | < ±2% fsd | < ±1% fsd | < ±2% fsd | < ±10% fsd | |||
Weight | 190g | 190g | 185g | 185g | 185g | 185g | 205g | 185g | 185g | 185g | 185g |
Construc. Materials | Polyphenylene Sulfide (PPS) | ||||||||||
I.P. (NEMA) Rating | IP65 (standard), IP67 (with weather protection fitted) | ||||||||||
Certification Details (ATEX/ IECEx) | Ex II 2 GD Ex d IIC Gb T85ºC Tamb -40ºC to +55ºC | Ex II 2 GD Ex d ia IIC T4 Gb Tamb -40ºC to +65ºC | |||||||||
Ex II 2 GD Ex d IIC Gb T100ºC Tamb -40ºC to +70ºC | Ex tb IIIC A21 IP67 T135ºC Db | ||||||||||
Ex II 2 GD Ex d IIC Gb T135ºC Tamb -40ºC to +80ºC | Baseefa08ATEX0263X | ||||||||||
Ex tb IIIC A21 Db IP67 | IECEx BAS08.0070X | ||||||||||
Baseefa08ATEX0265X | |||||||||||
IECEx BAS08.0068X | |||||||||||
Other Certifications | Russian: | GOST | |||||||||