Description
Honeywell MIDAS Sensor Cartridge for Carbon Dioxide ECC cartridge 0.15-2.00 % v/v
Several Honeywell MIDAS® cartridges can be configured to read a number of different gases based on a common chemical family. Actual details for the sensor detection range are included in the sensor cartridge specification sheet and the MIDAS® technical manual. MIDAS® sensor cartridges can be ordered as part of a Kit with the universal MIDAS® transmitter, or as a replacement part. The detectable range is listed. With the exception of O2 (where ambient concentration is typically 20.9%), the sensors will display zero when no gas is present. All Cartridges with the format MIDAS-E-XXX are supplied with a 2 year (24 month) warranty.
Gas Measured | Carbon Dioxide (CO2) | |
---|---|---|
Cartridge Part Number | MIDAS-E-CO2 | 2 year extended warranty |
Sensor Technology | 3 electrode electrochemical cell | |
Measuring Range (ppm) | CO2 | 0 – 2% v/v |
Minimum Alarm 1 Set Point | 2500 ppm | |
Repeatability | < ± 0.06% v/v | |
Linearity | < ± 15% of measured value | |
Response Time t62.5 | < 70 seconds | |
Sensor Cartridge Life Expectancy | ≥ 24 months under typical application conditions | |
Operating Temperature | 0°C to +40°C (32°F to 104°F) | |
Effect of Temperature | < 0.0039% vol / °C (0°C to 20°C) | |
Zero | < ± 0.0053% vol /°C (20°C to 40°C) | |
Sensitivity | < 1.9% of measured value / °C (0°C to 20°C) | |
< ± 0.9% of measured value / °C (20°C to 40°C) | ||
Operating Humidity (continuous) | 15 – 95% rH | |
Effect of Humidity | Zero | No effect |
Sensitivity | < ± 1% of measured value / % rH | |
Operating Pressure | 90 – 110kPa | |
Effect of Position | No effect in typical application | |
Long Term Drift | Zero | < +0.3% vol / year |
Sensitivity | < 30% of measured value / 1 year | |
Calibration Gas | Carbon Dioxide (CO2) | |
Challenge Gas (Bump Test) | Carbon Dioxide (CO2) | |
Warm Up Time | < 20 minutes | |
Storage Temperature | +5°C to +25°C (+41°F to +77°F) |
Each Midas sensor is potentially cross sensitive to other gases and this may cause a gas reading when exposed to other gases than those originally designated. The table below presents typical readings that will be observed when a new sensor cartridge is exposed to the cross sensitive gas (or a mixture of gases containing the cross sensitive species).
Gas / Vapor | Chemical Formula | Concentration applied (ppm) | Reading (ppm CO2) |
---|---|---|---|
Carbon Monoxide | CO | 500 | < 0.05 |
Chlorine | Cl2 | 2 | < 0.05 |
Hydrogen | H2 | 1000 | < 0.05 |
Hydrogen Sulfide | H2S | 75 | < 0.05 |
Methane | CH4 | 20000 | < 0.05 |
Nitrogen Dioxide | NO2 | 50 | < 0.05 |
Sulfur Dioxide | SO2 | 50 | < 0.05 |
Methanol | CH3OH | 1000 | < 0.05 |